Introduction
Since its beginnings on the 19th century, thyroid surgery has aimed to
resect thyroid diseases as well as to preserve the structures adjacent
to the thyroid gland, such as the parathyroid glands and the recurrent
laryngeal nerve (RLN), that are of utmost importance for maintenance of
quality of life. Postoperative RLN palsy prevention was achieved through
visual identification. Both Lahey and Riddell became advocates of
routine nerve dissection and visualization during thyroidectomy, and
this has become the gold standard procedure after several studies
confirmed that RLN integrity is protected by routine visual
identification.
Nonetheless, in clinical practice, RLN palsy can ensue even if the nerve is anatomically intact. This gap between anatomical and functional nerve preservation was solved with intraoperative neuromonitoring (IONM).
Before starting the procedure
The available devices convert the response of the laryngeal muscles to
recurrent laryngeal nerve stimulation into electromyographic signals
which can be observed visually and acoustically.
It is necessary to take into consideration the following components:
a. The stimulation circuit consists of a probe, a grounding electrode, and the connections to the interface.
b. The recording circuit consists of the recording electrodes fixed to
the endotracheal tube, the recording electrode ground and the
connections to the interface box.
It is important to underline that the dislocation of the recording electrodes after the endotracheal tube has become displaced is one of the main equipment-related problems. The right position of the endotracheal tube at the oral commissure should be marked on the endotracheal tube at the beginning of the case. The tracheal tube occupies its ideal position when the electrodes are located at the glottis and the endotracheal cuff is located at the subglottis.
IONM is sensitive to mechanical stimulation of the RLN and vagus nerve, and thus muscle relaxants that interfere with signal transmission at the neuromuscular junction may make it difficult to interpret the IONM.
How does it work
Stimulation of the vagus nerve and RLN during thyroid surgery elicits an
EMG response which reflects the function of the nerve. The minimum
current to produce a twitch in the posterior laryngeal wall ranges from
0.36 to 0.39 mA. The EMG amplitude of the IONM signal reach climax at
0.8 – 1.0 mA. Supramaximum stimulation with 1.0 to 2.0 mA enables
optimum functional assessment of both nerves during the procedure.
Interpretation
IONM is employed for confirmation of RLN function during thyroid surgery
as a loss of audiovisual response may signify RLN dysfunction or palsy
even if the nerve is anatomically intact.
Loss of audiovisual response signal is defined as a signal loss in a patient who initially demonstrated normal vocal cord function and strong EMG waveforms > 100 µV.
Loss of IONM in the face of an intact RLN may arise from:
a. Dislocation of the tracheal tube.
b. Administration of muscle relaxants.
c. Blood, fluid or decreased temperature around the RLN that impede transmission of the electric stimuli.
d. True loss of nerve function resulting in a transient or permanent nerve palsy.
Indications
IONM allows the surgeon to detect nerve injuries during neck surgery,
and not after it. This early diagnosis is essential to enable the
surgeon to react on the spot, by administering intraoperative steroids
or calcium channel blockers to improve nerve regeneration or by changing
the surgical strategy (postponing the contralateral thyroid resection).
RLN stimulation allows the mapping of the RLN variants that are difficult to detect. In particular, atypical RLN variants entail a higher risk of nerve injury. Early recognition of aberrant nerves may help decrease surgical morbidity secondary to atypical RLN variants.
Literature review
After several randomized clinical trials comparing IONM vs visual nerve
identification, a Cochrane review recently published (January 2019)
concluded that there is no conclusive evidence of the superiority or
inferiority of IONM over visual nerve identification.
Conclusions
Although there is no firm evidence concerning the superiority of IONM
over visual nerve identification, the advantages of IONM, such as
intraoperative detection of a nerve injury and prevention of damage by
detecting atypical RLN variants, may make IONM a useful tool in neck
surgery.